Plastic Medical Oxygen Sensor 3 Pin Envitec O2 Sensor 16mV Signal Output
Measurement
range
|
0 % ... 100 % oxygen (at atmospheric pressure) |
Nominal
sensor
lifetime
|
≥ 1 000 000 % volume oxygen hours |
Output
in
ambient
air
|
13
mV
...
16
mV
|
Electrical
interface
|
3.5
mm
mono
phone
Jack
|
Accuracy
|
meets
ISO
80601-2-55
requirements
|
Repeatability
|
<
1
%
volume
O2
at
constant
temperature
and
pressure
|
Linearity
error
|
<
3
%
relative
|
Response
time
|
<
12
s
to
90
%
of
final
value
|
Zero
offset
voltage
|
< 200
µV
in
100
%
nitrogen,
applied
for
5
min
|
Cross
interference
|
meets ISO 80601-2-55 requirements |
Influence
of
humidity
|
-0.03 % rel. per % RH at 25 °C |
Pressure range | 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2) |
Influence of pressure | proportional to change in oxygen partial pressure |
Influence of mechanical shock | < 1 % relative after a fall from 1 m |
Operating temperature | 0 °C ... +50 °C |
Temperature
compensation
|
built-in NTC compensation |
Effect
of
temperature
compensation
(steady
state)
|
between
+25
°C
and
+40
°C:
3
%
relative
error
between
0
°C
and
+50
°C:
8
%
relative
error
|
Operating
humidity
|
0 % ... 99 % RH non-condensing |
Long
term
output
drift
|
1
%
volume
oxygen
per
month
typically
<
-15
%
relative
over
lifetime
|
Storage
temperature
|
-20 °C ... +50 °C |
Recommended
storage
|
+5 °C … +15 °C |
Recommended
load
|
≥ 10 kOhms |
Warm-up time | < 30 minutes, after replacement of sensor |
Weight | approximately 28 grams |
1.
In-Warranty
Period:
we
will
give
the
replacement
for
all
quality
problems
in
time
under
normal
conditions
of
use.
2.
If
the
new
items
can
not
work
with
your
devices,
we
will
send
new
replacement
or
refund
in
time