Price | Negotiable |
MOQ | 1PCS |
Delivery Time | 5-8 work days |
Brand | KACISE |
Place of Origin | CHINA |
Certification | CE |
Model Number | KGD-412 |
Packaging Details | each unit has individual box and all boxes are packed in standard packages or customers requests available |
Payment Terms | L/C,D/A,D/P,T/T,Western Union,MoneyGram |
Supply Ability | 1000 Piece/Pieces per Week negotiable |
Place of Origin | CHINA | Detection Range | 0 - 150mmHg |
Packaging Details | each unit has individual box and all boxes are packed in standard packages or customers requests available | Input Voltage | 5.00V (±5%) |
Average Current | <120mA | Model Number | KGD-412 |
Supply Ability | 1000 Piece/Pieces per Week negotiable | Certification | CE |
Brand Name | KACISE | Payment Terms | L/C,D/A,D/P,T/T,Western Union,MoneyGram |
Data Output | CO₂ concentration (mmHg), end - tidal CO₂, inhaled CO₂, respiratory rate | Delivery Time | 5-8 work days |
Minimum Order Quantity | 1PCS | Detection Principle | Non - Dispersive Infrared Technology (NDIR) |
Resolution | 0.1mmHg |
Attribute | Value |
---|---|
Detection Principle | Non - Dispersive Infrared Technology (NDIR) |
Data Output | CO₂ concentration (mmHg), end - tidal CO₂, inhaled CO₂, respiratory rate |
Detection Range | 0 - 150mmHg |
Resolution | 0.1mmHg |
Input Voltage | 5.00V (±5%) |
Average Current | <120mA |
Mainstream End - Tidal Carbon Dioxide Sensor KGD - 412 for Measuring End - Tidal CO₂ Concentration
The KACISE end - tidal ETCO₂ sensor KGD - 412 is a mainstream ETCO₂ sensor module designed based on the non - dispersive infrared detection principle, used for measuring the end - tidal CO₂ concentration. The KGD - 412 can be applied to monitors, medical ventilators, and anesthesia machines. It has the advantages of accurate and reliable readings, perfect waveforms, compact appearance, and flexible installation.
Items | Details |
---|---|
Detection Principle | Non - Dispersive Infrared Technology (NDIR) |
Data Output | CO₂ concentration (mmHg), end - tidal CO₂, inhaled CO₂, respiratory rate |
Detection Range | 0 - 150mmHg |
Resolution | 0.1mmHg |
Detection Accuracy |
0
-
40mmHg:
±2mmHg 41 - 70mmHg: True value ±5% 71 - 100mmHg: True value ±8% 100 - 150mmHg: True value ±10% |
Respiratory Rate Range | 0 - 150BPM |
Respiratory Rate Accuracy | ±1 breath |
Stability |
Short
-
term
drift:
not
exceeding
0.8mmHg
(4H) Long - term drift: maintaining accuracy specifications within 120H |
Input Voltage | 5.00V (±5%) |
Average Current | <120mA |
Peak Current | <250mA |
Initialization Time | Display within 15s, full specification within 2min (25°C) |
Compensation | Air pressure: 400 mmHg - 850 mmHg |
Working Temperature and Humidity |
Operation:
0°C
-
45°C,
10
-
90%
RH,
non
-
condensing Storage: -40°C - 70°C, < 90% RH, non - condensing |
Data Communication | RS232 |
Serial Port | 19200 - N - 8 - 1 |
Dimensions | 22 mm * 23 mm * 36mm |