Place of Origin | Japan | Packaging Details | Carton |
Applicable Standard | C22.2 No.0-10 (R2015) | Model Number | FLXA21-D-P-D-AB-C1-NN-A-N-LA-N-NN |
Supply Ability | 100 | Certification | - |
Control Drawing | Refer to App. 5.2 | Brand Name | Yokogawa |
Payment Terms | L/C, D/A, D/P, T/T, Western Union, MoneyGram | Price | USD450/PC |
Delivery Time | 2-3WEEKS | Marking/Rating: | Ex ia IIC T4 Ga |
Minimum Order Quantity | 1 | Humidity | 10 to 90% RH at 40ºC (Non-condensing) |
Warrany | 1 year |
The FLXA202/FLXA21 is weatherproof and can be installed both inside and outside. It should, however, be installed as close as possible to the sensor to avoid long cable runs between the instrument and sensor. When the pH sensor is used, the cable length including the sensor cable should not exceed 20 meters (65.6 feet); 60 meters (197 feet) when using BA10 extension box and WF10 cable. For a conductivity sensor the cable run should not exceed 60 meters (197 feet). For dissolved oxygen the sensor cable run should not exceed 30 meters (100 feet). For SENCOM pH/ORP the sensor cable run should not exceed 20 meters (65.6 feet). Select an installation site where the ambient temperature and humidity are within the limits of the instrument specifications (see chapter 2). If the instrument is installed outside and exposed to direct sunlight, a sun shade hood should be used. Ambient Operating Temperature: -20 to +55 ºC Storage Temperature: -30 to +70 ºC Humidity: 10 to 90% RH at 40ºC (Non-condensing) Select an installation site that meets the following conditions. • Mechanical vibrations and shocks are negligible • No relay switch and power switch are installed close to the converter • There is space for cable connection beneath the cable glands • Not exposed to direct sunlight or severe weather conditions • Maintenance is possible • No corrosive atmosphere • Water Protection: IP66, NEMAType 4X (USA), Type 4X(Canada
The
method
to
connect
the
grounding
terminal
is
different
for
the
FLXA202
and
FLXA21.
For
the
safety
of
the
user
and
to
protect
the
instrument
against
interference,
the
converter
housing
must
always
be
connected
to
a
class
D
ground
(a
ground
resistance
of
100
ohm
or
less).
l
Grounding
the
FLXA202
housing
Use
the
terminal
outside
the
converter
housing
for
grounding
(Figure
2.9A).
Use
a
ring
terminal.
When
connecting
the
converter
to
a
distributor,
connect
the
shield
of
a
double-core
shielded
cable
to
the
terminal
on
the
terminal
block
in
the
converter.
Do
not
connect
the
shield
to
the
ground
terminal
in
the
distributor.
l
Grounding
the
FLXA21
housing
Connect
a
single
separate
cable
for
grounding
to
the
terminal
block
in
the
housing
(Figure
2.9B).
Use
a
cable
with
an
outer
diameter
of
3.4
to
7
mm.
Since
the
cable
is
very
thin,
set
the
sleeve
inside
the
standard
cap
of
the
cable
gland
and
pass
the
cable
through
it.
Sleeves
are
supplied
with
cable
glands
(Figure
2.6).
When
connecting
the
converter
to
a
distributor,
connect
the
shield
of
a
double-core
shielded
cable
to
the
terminal
on
the
terminal
block
in
the
converter.
Do
not
connect
the
shield
to
the
ground
terminal
in
the
distributor
Storage Temperature | -30 to +70 ºC |
Ambient Operating Temperature | -20 to +55 ºC |
Water Protection | IP66 |